Code Alarm CR3 Manuel de service Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 66
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
MIQ/CR3; DIQ/CR3 Settings
4 - 3
ba76032e01 01/2012
4.1.2 Limit indicator
With a limit indicator, a relay switches when a specified limiting value is
exceeded or undercut.
Limit indicators can be used in the following way:
Monitoring a limiting value using a relay:
when a limiting value (upper or lower limiting value) is exceeded or
undercut, a relay switches. The Open or Close relay actions are
possible in each case (see page 4-4).
Monitoring two limiting values using two relays:
If the upper limiting value is exceeded or undercut, a relay switches,
and if the lower limiting value is exceeded or undercut, another relay
switches. The Open or Close relay actions are possible in each case
(see page 4-4).
Note
If the simple monitoring function (Open, Close) with one or two relays
is not sufficient, use proportional output (see section 4.1.3).
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 65 66

Commentaires sur ces manuels

Pas de commentaire