MIQ/CR3; DIQ/CR3 Settings
4 - 37
ba76032e01 01/2012
4.6.3 Fixed current value
Function With the Fixed current value function, you can test the operativeness of
the instruments connected to the outputs: output different current
values to the output and, while doing so, check the behavior of the
connected instrument.
Note
The settings for other functions in the Current output menu, as for
example PID controller and Recorder, are retained while the Fixed
current value is carried out.
Setting Selection/Values Explanation
Current output Fixed current
value
Using Save and quit, the
nominal amperage that
was input as Inom is
output.
Inom 0 ... 20 mA The nominal amperage
that is output.