Settings MIQ/CR3; DIQ/CR3
4 - 42
ba76032e01 01/2012
4.9.1 Switching on the maintenance condition
4.9.2 Switching off the maintenance condition
1 Call up the measured value display with m.
2 Select the sensor you want to switch on the maintenance
condition for with
d. The measured values of the sensor do
not flash.
3 Open the 300 - Display/Options menu with g.
4 Using
d, highlight the menu item, Switch maintenance
condition on/off or Maintenance of sensor S0x and confirm with
g. A window that informs you about the maintenance
condition opens up.
5 Confirm Continue with g.
The selected sensor is in the maintenance condition. Linked
outputs are frozen.
6 Call up the measured value display with
m.
The measured values of the sensor flash.
1 Call up the measured value display with m.
2 Select the sensor you want to switch off the maintenance
condition for with
d. The measured values of the sensor
flash.
3 Open the 300 - Display/Options menu with g.
4 Using
d, highlight the menu item, Switch maintenance
condition on/off or Maintenance of sensor S0x and confirm with
g. A window that informs you about the maintenance
condition opens up.
5 Confirm Continue with g.
The maintenance condition of the selected sensor is switched
off. Linked outputs are released.
6 Call up the measured value display with
m.
The measured values of the sensor do not flash.