Troubleshooting
Problem Possible Cause Possible Solution
Model 7500 will
not activate.
The unit has no power. Plug in the AC adapter.
Model 7500 will
not operate on
batteries.
The battery pack is not
charged.
Plug in the Model 7500 AC Adapter
to charge the battery pack.
The battery pack is
inoperable.
Contact NONIN Customer Support
for repair or replacement.
Unable to obtain a
green pulse
display on the
bargraph.
The patient pulse
strength is
indiscernable or
perfused poorly.
Reposition the finger or insert a
different finger, and keep the
sensor motionless for at least 10
seconds.
Warm the patient’s finger by
rubbing or covering with a blanket.
Position the sensor at a different
site.
NOTE: In some
instances, patient
perfusion may be
inadequate for
pulse detection.
Circulation is reduced
because of excess
pressure on the
sensor (between the
sensor and a hard
surface) after inserting
finger.
Allow the hand to rest comfortably
without squeezing or pressing the
sensor on a hard surface.
The finger is cold. Warm the patient’s finger by
rubbing or covering with a blanket.
Position the sensor at a different
site.